In certain situations, the electronic loads are used in series for some special purposes. 2.1 Soft ON-OFF & Over current protection In CC mode, when the set current value of the load is bigger than the short-circuit current value, the electronic load has the following two functions:
1) Soft On-OFF Function. Users can control the circuit on/off by the keys at the front panel, and the hardware trigger port and RS232 interface at the rear panel.
2.2 A programmable high-power CC source 2.6 A milliohm meter When the remotes sensing function of the load is enabled, and the remote input of the load is connected to the both ends of the load resistor (RL), the equivalent resistance displayed in real-time by the load is the actual resistance of the RL. This application can serve as a 4-wire milliohm meter, used to test the resistance. Please note that power displayed in real-time by the load at the moment is not the real power of the load. If users want to conduct such kind of application, please ensure that actual loading power is less than the load’s rated power. Otherwise the load might be damaged. |